Key result
Zephyr chest bands track RR intervals during mental stress within ~9 ms of medical ECGs.
Why the study?
Wearable sensors for HRV-based mental stress assessment require minimal intrusiveness and inconspicuousness, but the accuracy trade-off of less intrusive chest band sensors compared to medical ECG devices was not quantified.
Does the Zephyr HxM chest band heart rate monitor accurately measure HRV metrics compared to a certified medical ECG device in young healthy subjects under stress?
Observational (n=14)
Does the Zephyr HxM chest band heart rate monitor accurately measure HRV metrics compared to a certified medical ECG device in young healthy subjects under stress?
The Zephyr HxM chest band sensor provides accurate RR interval and HRV measurements comparable to a medical-grade ECG during mental stress testing.
Answers are grounded in this paper and cite their sources.
May support wearable RR monitoring in stress research; leaves open validation in clinical populations.
Heart rate variability (HRV) analysis has been used as a quantitative marker of the autonomous nervous system activity to measure mental stress. Wearable sensors have been emerging as a solution to collect HRV data for stress assessment in a real context, however such studies raise additional requirements. The wearable system must be minimally obtrusive to allow the subjects to perform their tasks without interference, and inconspicuous to avoid the anxiety associated with wearing medical devices in public. The purpose of this study was to quantify the accuracy trade-off in the use of a chest band heart rate sensor that is less intrusive and less costly than a wearable electrocardiogram (ECG). The HRV metrics extracted from a chest band heart rate monitor, Zephyr HxM TM (Zph TM ), were compared with those extracted from an ECG certified medical device, Vital Jacket TM (VJ TM ). The two systems were worn simultaneously under laboratory conditions by a population of 14 young and healthy subjects, aged 20 to 26 years, under the stress induced by the Trier Social Stress Test (TSST) procedure. The results showed a mean difference between RR intervals of 9 ms; a root-mean square error (RMSE) of less than 8% and a Pearson’s correlation higher than 0.946, considering all TSST phases. In the HRV analysis, the average of all normal intervals (AVNN) showed errors less than 2% between the two systems with a correlation higher than 0.99 for all TSST phases. We thus conclude that the used chest band sensor represents an alternative to the current wearable medical devices to monitor RR intervals, and could be used for mental stress monitoring similar to the TSST protocol.
No takes yet. Share an insight, caveat, or question.
Pereira et al. (2017) conducted an observational in Healthy subjects (n=14). Chest band heart rate monitor (Zephyr HxM) vs. ECG certified medical device (Vital Jacket) was evaluated on RR intervals and HRV metrics (AVNN) (MD 9 ms). A chest band heart rate monitor (Zephyr HxM) showed high agreement with a medical ECG device for RR intervals (mean difference 9 ms, r > 0.946) during induced mental stress.