Key result
WRAPSODY endoprosthesis achieves 100% 30-day target lesion primary patency in arteriovenous access stenoses.
Why the study?
The safety and effectiveness of the Cell-impermeable Endoprosthesis for treating arteriovenous access circuit stenosis in haemodialysis patients required clinical evaluation.
Does a cell-impermeable endoprosthesis improve patency and safety in haemodialysis patients with access circuit stenosis?
Cohort (n=46)
Yes
Does a cell-impermeable endoprosthesis improve patency and safety in haemodialysis patients with access circuit stenosis?
The cell-impermeable endoprosthesis demonstrated high procedural success and 12-month target lesion primary patency with a favorable safety profile in hemodialysis patients with access circuit stenosis.
Supports early consideration of cell-impermeable endoprosthesis for AV access stenosis; leaves open confirmation of durability in randomized trials.
PURPOSE: Cell-impermeable Endoprosthesis (Merit Medical Systems, Inc.) in the treatment of arteriovenous fistula and arteriovenous graft access circuit stenosis. MATERIALS AND METHODS: Investigators conducted a prospective analysis of 46 patients with access circuit stenosis from three centres. Treatment sites included the peripheral outflow veins (e.g. cephalic arch, basilic vein swing point; 16 fistula and 10 graft patients); the graft-vein anastomosis (9 patients); and the central veins (up to, but not including the SVC; 11 patients). Primary outcome measures included 30-day freedom from access circuit-related safety events and 30-day target lesion primary patency. Secondary outcome measures included procedural success; device- and procedure-related adverse events; target lesion primary patency; access circuit primary patency; and secondary patency. In-person follow-up was scheduled at 1, 3, 6, and 12 months. An independent data monitoring/clinical event committee adjudicated all reinterventions and device/procedure-relatedness for adverse events. RESULTS: All initial procedures were successful. All but one patient was free from safety events through the first 30 days (97.8% (45/46)). This event was not device-related. Over the remainder of the study, one adverse event was adjudicated as possibly device-related. Six- and 12-month target lesion primary patency rates were 97.7% (42/43) and 84.6.% (33/39), respectively. Six- and 12-month access circuit primary patency rates were 84.4% (38/45) and 65.9% (29/44), respectively. CONCLUSION: Results suggest that the study device is safe and effective for treatment of stenoses in the peripheral and central veins of arteriovenous access circuits. LEVEL OF EVIDENCE: Level 2b, cohort study.
No takes yet. Share an insight, caveat, or question.
Gilbert et al. (2021) conducted a cohort in Arteriovenous access circuit stenosis (n=46). WRAPSODY Cell-impermeable Endoprosthesis was evaluated on 30-day target lesion primary patency (TLPP). The WRAPSODY Cell-impermeable Endoprosthesis safely and effectively treated arteriovenous access circuit stenoses, achieving 100% target lesion primary patency at 30 days.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: