Sixty-nine patients with a Greenfield inferior vena cava (IVC) filter in place for 1-9 years were evaluated with supine abdominal radiographs. In 22 patients (32%) the filter span diameter had increased by 3-11 mm, and in six patients (9%) it had decreased by 3-18 mm. Twenty patients (29%) had caudad migration of 3-18 mm, and four (6%) had cephalad migration. In five patients (7%) the filter was at an angle of more than 16 degrees from the vertical. A decrease in filter span of 7 mm or more may result in IVC occlusion, whereas an increase of 6 mm or more may result in caval wall penetration. Caudad migration may be caused by a fibrous reaction around the struts which slowly draws the filter caudally. A change in filter angle may result from displacement of a strut into the right renal vein as well as from physiologic changes. The abdominal radiograph is a valuable method for follow-up study of patients with Greenfield filters if the filter position, span, migration, and angle are noted in the report.
No takes yet. Share an insight, caveat, or question.
Messmer et al. (1985) studied this question.