We demonstrate arrays of cantilevers with different lengths, fabricated by focused ion beam milling. The arrays of oscillators generate a spectrum of different resonant frequencies, where each frequency correlates to the corresponding individual cantilever. The frequency response from all the cantilevers is collected from a single measurement under the same environment and conditions for the entire array. The mass response of the system generated the same Δ f / f 0 for the cantilevers, within 0.1% accuracy. We denote the method MFSAC: multi-frequency signal analysis from an array of cantilevers. The simultaneous detection of several frequencies in one spectrum has great benefits in mass sensor applications, offering the possibility for true label-free detection.