Answers are grounded in this paper and cite their sources.
BACKGROUND: Validated assays are essential to generate data with defined specificity, consistency, and reliability. Although the process of validation is required for applying immunoassays in the context of clinical studies, reports on systematic validation of in vitro T cell assays are scarce so far. We recently validated our HLA-peptide multimer staining assay in a systematic manner so as to qualify the method for monitoring antigen-specific T cell responses after immunotherapy. METHODS: Parameters of the assay, specificity, precision, linearity, sensitivity, and robustness were assessed systematically. Experiments were designed to address specifically each parameter and are detailed. CONCLUSIONS: We succeeded in validating the HLA-multimer staining assay in a systematic manner. Additionally, we propose a technical framework and recommendations that can be applied for validating other T cell assessment methods.
No takes yet. Share an insight, caveat, or question.
Chandran et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: