Answers are grounded in this paper and cite their sources.
Low pressure measurements by Bayard-Alpert gauges are limited by x-ray photoelectric currents at the ion collector, equivalent to a pressure reading of about 1×10−11 Torr. Through efforts to reduce this x-ray current, the Bayard-Alpert structure has been modified, whereby a beam of ions is extracted from the grid and bent around a corner to a remote collector. By this geometry the collector is shielded from the x-ray source at the grid, and the photoelectric current is reduced by a factor of 100, without incurring a loss in gauge sensitivity. Bending of the ion beam is accomplished by a cylindrical, electrostatic field, in which the ion trajectories are independent of e/m. 100% modulation of the ion beam intensity is achieved by turning off the bending field, thereby permitting pressure measurements as low as 1×10−14 Torr. A total pressure of 5×10−13 Torr, nitrogen equivalent, has been measured in a stainless steel vacuum system employing ion sputtering and titanium sublimation as the means of evacuation.
No takes yet. Share an insight, caveat, or question.
Helmer et al. (1966) studied this question.