Key result
Novel hydraulically steerable soft catheter achieves ~1.1-second response times with in vivo porcine visibility.
Why the study?
Steerable catheters offer advantages over conventional catheters by enhancing control, stability, and accessibility, but improvements in flexibility and safety remain needed for vascular interventions.
A novel hydraulically steerable soft-tip catheter demonstrated favorable mechanical properties and successful in vivo validation in a porcine model, highlighting its potential for vascular interventions.
May enhance vascular access and reduce radiation; leaves open clinical validation before practice adoption.
Steerable catheters offer significant advantages over conventional catheters, including enhanced control, stability, and accessibility, which reduce operational complexity, fluoroscopy time, and radiation exposure, positioning them as a promising advancement for vascular interventional procedures. Herein, a novel steerable catheter is presented, featuring a hydraulically actuated, soft, steerable tip that allows for real‐time visualization in X‐ray imaging. To optimize performance, several silicone materials were evaluated for their mechanical properties, resulting in a soft tip design with a diameter of 2.6 mm. The tip incorporates an internal tool channel and supports a large bending angle of 180°. The tip demonstrates an average response time of 1.141 s (±0.750 s), a maximum output force of 0.145 N (±0.001 N), and a maximum radial expansion of 1.121 (±0.006). A steering kinematic model of the catheter tip is developed to simulate its movement. The catheter tip's real‐time shape and position information are obtained through intelligent segmentation and neighborhood‐based endpoint detection methods, assisting the surgeon during superselective procedures. The catheter's visibility and flexibility are validated in a live porcine model, demonstrating its potential for future use in interventional procedures.
No takes yet. Share an insight, caveat, or question.
Kang et al. (2025) studied Vascular interventions. Hydraulically actuated, soft, steerable catheter tip was evaluated on Mechanical properties (response time, output force, radial expansion) and visibility/flexibility in a porcine model. A novel hydraulically steerable catheter with a soft tip demonstrated an average response time of 1.141 s and a maximum output force of 0.145 N, with validated visibility in a live porcine model.