The analytical performance of high-resolution multiple-collector inductively coupled plasma mass spectrometry (HR-MC-ICPMS), employing amplifiers with a 1012 ohm resistor, was examined. HR-MC-ICPMS showed 50 times higher detection efficiency and 100 times better precision for sulfur and 34S/32S than the single-collector high-resolution sector field type ICPMS (HR-ICP-SFMS). The 3σ detection limits for 32S and 34S were 0.2 and 6 ng mL−1, respectively. No matrix effects were observed, for up to 3000 μg g−1 of total dissolved solids (TDS), within 0.6% error. Thus when the isotope dilution (ID) technique is utilized, this method allows analyses of sulfur concentration of seven times smaller, with six times better precision, than the HR-ICP-SFMS methods. To demonstrate the applicability of this ID-HR-MC-ICPMS method, bulk sulfur concentrations in bovine serum albumin, human serum, eight silicate reference materials (JB-2, JB-3, JA-2, JA-3, JP-1, BCR-2, BHVO-1 and AGV-1), and four carbonaceous chondrites (Murchison, Allende, Dal al Gani 521 and 194) were measured.
No takes yet. Share an insight, caveat, or question.
Makishima et al. (2012) studied this question.