This is the authors' abstract. We don't add key points for this paper.
We present and demonstrate a miniature contact force sensor based on a 30-μm diameter microfiber Bragg grating. The microfiber Bragg grating packaged in a conforming elastomer material features an extremely high sensitivity of 1.37 mN and excellent linearity up to a contact force of 0.65 N. The performances were experimentally characterized and meet all the essential specifications as minimally invasive surgery devices for catheters' applications. The novel force sensor is easy to fabricate and thus potentially low cost.
No takes yet. Share an insight, caveat, or question.
Chung et al. (2012) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: