Key result
Digitized paper AI-ECGs show high agreement with native digital ECGs for detecting low LVEF.
Why the study?
The deep learning model for detecting left ventricular dysfunction depends on digital ECG data, limiting validation and scalability to institutions with paper ECGs.
Do deep learning models using digitized paper ECGs agree with native digital ECGs for detecting low ejection fraction in patients with varying ECG features?
Cross-Sectional (n=10)
Do deep learning models using digitized paper ECGs agree with native digital ECGs for detecting low ejection fraction in patients with varying ECG features?
Effect estimate: ICC 0.95 for Model A, 0.58 for Model B (95% CI 0.82-0.99 for Model A, 0.00-0.87 for Model B)
Deep learning models can achieve high agreement and diagnostic performance for detecting low ejection fraction using digitized paper-based ECGs, expanding the potential utility of archived clinical datasets.
Enables DL models on digitized paper ECGs for low EF detection; hypothesis-generating and requires prospective validation before clinical use.
Recently, a deep learning model was developed and validated for detecting left ventricular dysfunction based on a standard 12-lead ECG. However, this model largely depends on the availability of digital ECG data: 10 seconds for all 12 leads sampled at 500 hertz stored as a numeric array. This limits the ability to validate or scale this technology to institutions that store ECGs as PDF or image files (“paper” ECGs). Methods do exist to create digital signals from the archived paper copies of the ECGs. The primary objective of this study was to evaluate how well the AI-ECG model output obtained using digitized paper ECGs agreed with the predictions from the native digital ECGs for the detection of low ejection fraction. To address this objective, deep learning models that utilizes digitized data from a 12-lead ECG snapshot were needed. Two models were evaluated, Model A using data from a single lead with full 10-second recording (lead II) only and Model B using data from 3 leads with 10-second recordings (leads II, V1 and V5) in addition to 9 leads with partial (2.5-second) recordings. In a test sample of 10 patients with varying ECG features, Models A and B obtained intraclass correlation coefficients of 0.95 (95% CI: 0.82 to 0.99) and 0.58 (95% CI: 0.00 to 0.87). In an exploratory examination of model diagnostic performance to detect low ejection fraction, Model A achieved and AUC of 0.71 while Model B achieved an AUC of 0.91. Our study demonstrates an agreement between deep learning model predictions obtained from digitized paper-based ECGs and native digital ECGs and provides some insight into potential expandability of ECG-based deep learning models including the importance of captured duration (10-second vs. 2-5-second recordings) and ECG vectors (precordial leads vs. limb leads).
No takes yet. Share an insight, caveat, or question.
Adedinsewo et al. (2022) conducted a cross-sectional in Left ventricular dysfunction (n=10). Digitized paper ECGs vs. Native digital ECGs was evaluated on Agreement between AI-ECG model output using digitized paper ECGs and native digital ECGs for the detection of low ejection fraction (ICC 0.95 for Model A, 0.58 for Model B, 95% CI 0.82-0.99 for Model A, 0.00-0.87 for Model B). Deep learning models using digitized paper ECGs agreed with native digital ECGs for detecting low ejection fraction, achieving an intraclass correlation coefficient of 0.95 (95% CI 0.82-0.99) for Model A.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: