In the present study, an HPLC method was developed and validated for the determination of four sulphonamides: sulphadiazine (SDZ), sulphaquinoxaline (SQX), suplhamethazine (SMTH), and sulphadimethoxine (SDM). Two of them, SQX and SDM, were determined in cow's milk. The analytical column, a Kromasil, C18 5 µm, 250×4 mm analytical column, was operated at ambient temperature. The mobile phase, a mixture of 0.5% acetic acid as solvent A, CH3CN as solvent B, and CH3OH was delivered to the analytical column according to a gradient program. PDA detection was performed for the detection and confirmation of separated analytes with monitoring at 260 nm. Method validation was performed by means of intra‐day (n=5) and inter‐day accuracy and precision (n=5), sensitivity, and linearity. Limits of detection (LOD) and limits of quantification (LOQ) were 13 and 40 µg/kg, respectively. Solid phase extraction was applied to remove all matrix interference from milk samples after deproteinization with 8 M HCl. High extraction recoveries (>84%) were achieved using DSC‐18 cartridges with CH3OH‐0.5% CH3COOH as eluent. CCα values were 111.8 and 117.1 µg/kg for SDM and SQX, respectively, and CCβ values were 116.6 and 134.0 µg/kg, respectively. The method was applied to the analysis of twenty two milk samples from the local market. SQX was identified in seven of these samples.
No takes yet. Share an insight, caveat, or question.
Samanidou et al. (2008) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: