An electrochemiluminescence (ECL) inhibition method was developed as an indirect detection method for the determination of dopamine and epinephrine separated by capillary electrophoresis (CE). When the concentration of Ru(bpy)(3) (2+) was 50 muM diluted by 50 mM phosphate (pH 8.5) in the cell and 0.5 M tripropylamine (TPA) was added to the running buffer (10 mM phosphate, pH 9.0), an inhibition of ECL of the Ru(bpy)(3) (2+)/TPA system by the analytes was observed. Under the optimized conditions, the relative standard deviations of migration time and negative peak area were less than 1% and 3%, respectively, for 1 microM dopamine or 1 microM epinephrine (n = 10). Linear ranges of 0.1-10 microM for both analytes and the detection limits (signal-to-noise ratio S/N = 3) of 10 nM for dopamine and 30 nM for epinephrine were obtained.