Answers are grounded in this paper and cite their sources.
This paper presents a new needle path planning method for the insertion of rigid needles into deformable tissue. The needle insertion point, needle heading, and needle depth are optimized by minimizing the distance between a rigid needle and a number of targets in the tissue. The optimization method is based on iterative simulations performed using a tissue finite element model. At each iteration, the best 3-D line fitted to the displaced targets in the deformed tissue is used as a candidate for a new insertion line. First, this method is implemented in a prostate brachytherapy simulator under different boundary conditions to minimize the targeting error. It is shown that the optimization method converges in a few iterations and decreases the seed misplacement error to less than the needle diameter. Second, the efficacy of the optimization algorithm is verified by optimizing the insertion parameters for a brachytherapy needle before insertion into a prostate tissue phantom. The elastic properties of the phantom and the needle-tissue interaction parameters were identified in an independent experiment. The optimization algorithm is effective in decreasing the targeting error.
No takes yet. Share an insight, caveat, or question.
Dehghan et al. (2009) studied this question.