Answers are grounded in this paper and cite their sources.
A displacement monitor for use with mechanical stimulators with range of 2.5 mm, ±5-percent linearity, l-μm resolution, and less than 25-μs rise time can be built inexpensively using a solid-state interrupter module. Operational amplifier circuits are given for loudspeaker driver and monitor electronics.
No takes yet. Share an insight, caveat, or question.
Paul B. Brown (1974) studied this question.