Whole-body impulsive rotations were used to overcome several limitations associated with manual head impulse testing. A computer-controlled rotational chair delivered brief, whole-body, earth-vertical axis yaw impulsive rotations while eye movements were measured using video-oculography. Results from an unselected group of 20 patients with dizziness and a group of 22 control subjects indicated that the horizontal computerized rotational head impulse test (crHIT) is well-tolerated and provides an estimate of unidirectional vestibulo-ocular reflex gain comparable to results from caloric testing. This study demonstrates that the horizontal crHIT is a new assessment tool that overcomes many of the limitations of manual head impulse testing and provides a reliable laboratory-based measure of unilateral horizontal semicircular canal function.