Answers are grounded in this paper and cite their sources.
Laboratory study demonstrates accurate plasma creatinine detection using a portable photonic waveguide sensor, indicating potential for compact point-of-care diagnostic devices.
We demonstrate a portable, compact system to perform absorption-based enzymatic assays at a visible wavelength of 639 nm on a photonic waveguide-based sensor chip, suitable for lab-on-a-chip applications. The photonic design and fabrication of the sensor are described, and a detailed overview of the portable measurement system is presented. In this publication, we use an integrated photonic waveguide-based absorbance sensor to run a full enzymatic assay. An assay to detect creatinine in plasma is simultaneously performed on both the photonic sensor on the portable setup and on a commercial microplate reader for a clinically relevant creatinine concentration range. We observed a high correlation between the measured waveguide propagation loss and the optical density measurement from the plate reader and measured a limit-of-detection of 4.5 μM creatinine in the sensor well, covering the relevant clinical range for creatinine detection.
No takes yet. Share an insight, caveat, or question.
Neutens et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: