Key result
Custom-made inverted limb device successfully treats delayed Type IIIb post-EVAR endoleak percutaneously, avoiding open bypass.
Why the study?
Type III endoleak can be difficult to distinguish from Type I endoleak and standard bifurcated endografts may not be feasible for treatment depending on stent graft anatomy.
Case Report (n=1)
A custom-made device with an inverted contralateral limb can be a viable percutaneous treatment option for complex Type IIIb endoleaks presenting decades after initial EVAR.
May support percutaneous custom-device repair for rare late Type IIIb endoleaks; leaves open need for larger series on durability.
PURPOSE: Type III endoleak can be difficult to distinguish from Type I endoleak. Depending on the stent graft anatomy, the use of standard bifurcated endografts may not be technically feasible, and patients may have to be subject to an aorto-uni-iliac repair with femoral-femoral bypass or open surgery. CASE REPORT: We report a case of an 86-year-old male who had a Type IIIb endoleak 20 years post EVAR which was characterized on angiography to be from a hole close to the bifurcation limb origin. The initial Talent (Medtronic, Santa Rosa, California) device had a 50 mm main body common trunk, which was not amenable to treatment with standard devices. He was successfully treated with a custom-made device with an inverted contralateral limb. CONCLUSIONS: Our case highlights the need for lifelong surveillance post EVAR as endoleak may present decades post initial EVAR. It also demonstrates that many Type III endoleak which were otherwise deemed unsuitable for treatment with standard devices may potentially be treatable with custom-made device (CMD). This solution preserves a percutaneous option in a now older person which avoids surgical bypass. Further studies are required to establish the durability of this treatment and survey for recurrence.
No takes yet. Share an insight, caveat, or question.
Linn et al. (2022) conducted a case report in Type IIIb endoleak post EVAR (n=1). Custom-made device with an inverted contralateral limb was evaluated on Successful treatment of endoleak. A delayed Type IIIb endoleak 20 years post EVAR was successfully treated percutaneously using a custom-made device with an inverted contralateral limb, avoiding surgical bypass.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: