A high-performance liquid chromatographic method for the chiral separation of 1-benzyl-4 -[((5, 6-dimethoxy-1-indanon)-2-y1] methylpiperidine hydrochloride (E2020) and its plausible metabolites in plasma has been developed using an avidin column. The hydrophobicity (alkyl chain length) of a spacer, combining avidin to the silica support, affected significantly the retention and separation of E2020 enantiomers. A new column-switching technique using avidin columns for pretreatment and chiral separation enabled the direct injection analysis of plasma samples and the enrichment of E2020, which gave high-sensitivity. This occurs because plasma proteins are excluded from the avidin column, while E2020 enantiomers are retained on the column. This column-switching technique and Liquid chromatography-mass spectrometry were used to separate E2020 enantiomers from its plausible metabolites in plasma. This system enables to determine the level of E2020 enantiomers in plasma with high-selectivity and high-sensitivity without any pretreatments.
No takes yet. Share an insight, caveat, or question.
Oda et al. (1992) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: