Key result
Side-firing microwave array ablation achieves 100% atrial flutter noninducibility in sheep.
Why the study?
Long side-firing microwave arrays may enable rapid creation of long, continuous, transmural linear ablation lesions without intimate endocardial contact, but their efficacy in treating atrial flutter was untested.
Does percutaneous microwave ablation with a long side-firing antenna array achieve AFL noninducibility in an ovine atrial flutter model?
Does percutaneous microwave ablation with a long side-firing antenna array achieve AFL noninducibility in an ovine atrial flutter model?
High-efficiency side-firing microwave array ablation can rapidly create long linear and electrically intact lesions to successfully treat atrial flutter in an ovine model.
Answers are grounded in this paper and cite their sources.
Supports microwave ablation feasibility in ovine AFL model; hypothesis-generating for humans, requiring clinical trials before practice change.
INTRODUCTION: Long side-firing microwave (MW) arrays can deliver energy uniformly over its length without the need for intimate endocardial contact. We hypothesize that a novel 6 Fr 20 mm long percutaneous high-efficiency MW antenna array ablation catheter can rapidly create long, continuous, and transmural linear ablation lesions. METHODS AND RESULTS: Cavotricuspid isthmus (CTI)-dependent atrial flutter (AFL) was created in 11 sheep by a line of radiofrequency ablation lesions in the posterior right atrium (RA) linking the venae cavae. After 4-6 weeks recovery, CTI-dependent AFL was still inducible in all 11 sheep (cycle length 178 +/- 13 ms). MW ablation of the CTI at 100 W for 30 seconds was then performed with an endpoint of AFL noninducibility. AFL was not inducible in all 11 sheep after 4.3 +/- 3.3 MW applications (129 +/- 99 seconds). The last 6 animals needed fewer ablations (2.2 +/- 1.5) and 3 of these sheep required only a single ablation. Although conduction times from proximal coronary sinus to lateral RA and vice versa increased postablation (51 +/- 14 ms to 118 +/- 31 ms [P = 0.0002] and 60 +/- 13 ms to 119 +/- 28 ms [P = 0.0001], respectively), AFL was still inducible in 2 sheep and further ablation was needed to reach the endpoint. CONCLUSIONS: High-efficiency side-firing MW array ablation can rapidly create long linear and electrically intact lesions in an ovine AFL model. AFL noninducibility may be a more reliable indicator than CTI conduction times of an intact line of ablation in this animal model.
No takes yet. Share an insight, caveat, or question.
Lim et al. (2009) studied Cavotricuspid isthmus-dependent atrial flutter (n=11). Percutaneous high-efficiency microwave antenna array ablation catheter was evaluated on Atrial flutter noninducibility. High-efficiency side-firing microwave array ablation successfully achieved atrial flutter noninducibility in all 11 sheep after an average of 4.3 applications.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: