Key result
A fully automated method for estimating the peak frequency of the posterior dominant rhythm showed very high correlation with human ratings (rho = 0.94-0.95).
Why the study?
Does a fully automated method accurately estimate the peak frequency of the posterior dominant rhythm compared to human ratings in a retrospective EEG cohort?
Observational (n=9,197)
Blinded raters
Does a fully automated method accurately estimate the peak frequency of the posterior dominant rhythm compared to human ratings in a retrospective EEG cohort?
Effect estimate: rho = 0.94-0.95
A fully automated method can accurately estimate the peak frequency of the posterior dominant rhythm in EEGs, producing results highly correlated with human ratings.
OBJECTIVE: To develop and test a fully automated method for estimation of the peak frequency of the posterior dominant rhythm (PDR) in a large retrospective EEG cohort. METHODS: Thresholding was used to select suitable EEG data segments for spectral estimation for electrode O1 and O2. A random sample of 100 peak frequency estimates were blindly rated by two independent raters to validate the results of the automatic PDR peak frequency estimates. We investigated the relationship with age, sex and binary EEG classification. RESULTS: There were 9197 eligible EEGs which resulted in a total of 6104 PDR peak frequency estimates. The relationship between automatic estimates and age was found to be consistent with the literature. The correlation between human ratings and automatic scoring was very high, rho = 0.94-0.95. There was a sex difference of d = 0.33 emerging at puberty with females having a faster PDR peak frequency than males. CONCLUSIONS: Fully automatic PDR peak frequency estimation not dependent on annotated EEG produced results that are very close to human ratings. SIGNIFICANCE: PDR peak frequency can be automatically estimated. A compiled version of the algorithm is included as an app for independent use.
No takes yet. Share an insight, caveat, or question.
Zibrandtsen et al. (2020) conducted an observational in Hospital EEG cohort (n=9,197). Fully automated method for estimation of PDR peak frequency vs. Human ratings was evaluated on Correlation between human ratings and automatic scoring (rho = 0.94-0.95). A fully automated method for estimating the peak frequency of the posterior dominant rhythm showed very high correlation with human ratings (rho = 0.94-0.95).