Answers are grounded in this paper and cite their sources.
A method is presented to separate and detect six arsenic species by capillary electrophoresis (CE) interfaced to inductively coupled plasma mass spectrometry (ICP-MS). CE was used as a highly resolving separation system, whereas ICP-MS served as an element selective detector providing low detection limits. The special mode of operation included sample stacking and a differentiation of separation and detection. This provided separation and detection of six As species, uncharged and anionic, to be monitored within a single run. Detection limits were calculated according to IUPAC recommendation at 15 microg As/L for As (III), dimethyl arsinic acid (DA), monomethyl arsonic acid (MA) and As (V), or 65 microg As/L for arsenobetaine (AsB) and arsenocholine (AsC). Investigations were focused on possibly occurring interferences, e.g., ArCl+ interference at the monoisotope 75As. Finally, real samples from biomedical field (urine) and environmental field (sewage sludge) were analyzed.
No takes yet. Share an insight, caveat, or question.
Michalke et al. (1998) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: