Answers are grounded in this paper and cite their sources.
Purpose: Active matrix flat panel imagers (AMFPIs) with thin-film transistor arrays experience image quality degradation by electronic noise in low-dose radiography and fluoroscopy. One potential solution is to overcome electronic noise using avalanche gain in an amorphous selenium (a-Se) (HARP) photoconductor in indirect AMFPI. In this work, we aim to improve temporal performance of HARP using a novel composite hole blocking layer (HBL) structure and increase optical quantum efficiency (OQE) to CsI:Tl scintillators by tellurium (Te) doping. Approach: Two different HARP structures were fabricated: Composite HBL samples and Te-doped samples. Dark current and optical sensitivity measurements were performed on the composite HBL samples to evaluate avalanche gain and temporal performance. The OQE and temporal performance of the Te-doped samples were characterized by optical sensitivity measurements. A charge transport model was used to investigate the hole mobility and lifetime of the Te-doped samples in combination with time-of-flight measurements. Results: and achieved an avalanche gain of 16. Te-doped samples increased OQE from 0.018 to 0.43 for 532 nm light. The addition of Te resulted in 2.1% first-frame lag, attributed to hole trapping within the layer. Conclusions: The composite HBL and Te-doping can be utilized to improve upon the limitations of previously developed indirect HARP imagers, showing excellent temporal performance and increased OQE, respectively.