Answers are grounded in this paper and cite their sources.
Fifty-three of 55 consecutive elbow replacements for post-traumatic arthritis were followed for a minimum of two years (mean 6.3, range 2 to 14.4). The patients presented difficult management problems, having undergone an average of two previous operations per joint; 22 joints had suffered prior complications; 18 had less than 50 degrees of flexion and six were flail. One of three versions of the Coonrad prosthesis was employed in all. During the follow-up period, 10 patients underwent 14 revision procedures for aseptic loosening; 38 elbows are currently without progressive radiolucent lines. In two patients an elbow had to be resected, one for deep infection and the other for bone resorption following a foreign-body reaction to titanium. The current design of the Coonrad prosthesis offers a reliable option for the treatment of post-traumatic arthritis but should be used only in carefully selected patients over the age of 60 years.
No takes yet. Share an insight, caveat, or question.
Morrey et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: