While the qualitative effects of grid misalignment are known, we have quantified the effect of different degrees of grid misalignment on image contrast and patient exposure. Radiographs were made of a phantom consisting of five lead disks on top of a 15 cm block of lucite. Four 60 lines/cm grids, having grid ratios of 3:1, 4:1, 6:1, and 8:1 were used. When the tube was angled more than three degrees across the grid lines, the contrast improvement factor decreased substantially for all four grids, as much as 46% for an 8:1 grid with a 12 degrees misalignment. There was a concomitant decrease in film optical density, which if compensated for by an increase in patient exposure, would lead to a higher effective bucky factor. With the exception of the 3:1 grid, if the grid is misaligned by more than 6 degrees, higher signal-to-noise ratios can be attained by removing the grid and using the increased patient exposure to reduce noise.