Answers are grounded in this paper and cite their sources.
A particle sizing and counting instrument capable of working with concentrations as high as 105 particles/cm3 is described. The instrument measures inline in an optical, non-intrusive manner and has process control capabilities. Since testing high efficiency filters requires high input concentrations and since filter efficiency is a function of particle size, this instrument is well-suited for filter analysis. Results of several measurements on filters are shown, including how factors such as particle size, velocity, electrical charge and filter age influence the performance of a filter.
No takes yet. Share an insight, caveat, or question.
Umhauer et al. (1990) studied this question.