Answers are grounded in this paper and cite their sources.
For the calibration of low-energy neutral particle mass spectrometers a neutralization stage for a charged-particle beam was built. The incident positive ion beam is neutralized upon reflection at grazing incidence off a highly polished tungsten single crystal surface. With protons as primary species the device produced a neutral hydrogen beam in the energy range of 10 eV–1000 eV depending on incident ion-beam energy with a centre energy of 85% to 89% and an energy width of 11% to 15% FWHM of the incident ion energy. The angular spread of the neutral beam was limited by apertures to 4.6° × 18° FWHM. The overall transmission of the device was found to be between 0.005 and 0.013.
No takes yet. Share an insight, caveat, or question.
Wieser et al. (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: