A fiber-optic sensor is presented that is capable of measuring the particle displacement in high-intensity focused ultrasound (HIFU) fields. For this probe, a secondary calibration was performed, and the resulting complex frequency response is discussed. As a first practical application, the setup was used to measure the pressure in the field of a weakly focusing ultrasound transducer. The result is compared with that of a membrane hydrophone measurement. The feasibility of measurements in HIFU fields is demonstrated by means of measurements of the spatial distribution of the peak particle velocity within the focus of a HIFU transducer and of the dependence of the peak values on the acoustical power level.