Key result
Short-tip dilator enables safe mechanical aortic valve crossing for complex arch aneurysm repair.
Why the study?
A novel technique to cross a mechanical aortic valve prosthesis is needed to facilitate endovascular repair in complex aortic arch anatomy.
Case Report (n=1)
Passing a mechanical aortic valve prosthesis with a short tip dilator is a feasible technique to facilitate complex aortic arch endovascular branched repair.
May enable complex arch repair in patients with mechanical aortic valves; hypothesis-generating case report requiring prospective validation.
PURPOSE: To present a novel technique to successfully cross a mechanical aortic valve prosthesis. TECHNIQUE: A 55-year-old female patient with genetically verified Marfan syndrome presented with a 5-cm anastomotic aneurysm of the proximal aortic arch after previous ascending aortic replacement due to a type A aortic dissection in 2007. The patient also underwent mechanical aortic valve replacement in 1991. A 3-stage hybrid repair was planned. The first 2 steps included debranching of the supra-aortic vessels. In the third procedure, a custom-made double branched endovascular stent-graft with a short 35-mm introducer tip was implanted. The mechanical valve was passed with the tip of the dilator on the lateral site of the leaflet, without destructing the valve and with only mild symptoms of aortic insufficiency, as one leaflet continued to work. This allowed the implantation of the stent-graft directly distally of the coronary arteries. Postoperative computed tomography angiography showed no endoleaks and patent coronary and supra-aortic vessels. CONCLUSION: Passing a mechanical aortic valve prosthesis at the proper position is feasible and allows adequate endovascular treatment in complex arch anatomy. However, caution should be taken during positioning of the endovascular graft as the tip may potentially damage the valve prosthesis.
No takes yet. Share an insight, caveat, or question.
Konstantinou et al. (2021) conducted a case report in Anastomotic aneurysm of the proximal aortic arch (n=1). Passing a mechanical aortic valve with a short tip dilator was evaluated on Feasibility and success of crossing a mechanical aortic valve prosthesis. Passing a mechanical aortic valve prosthesis with a short tip dilator was feasible and allowed successful endovascular treatment of a complex arch aneurysm in a 55-year-old female.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: