This paper presents a fully functionally integrated 1024-channel mixed-mode and mixed-voltage system-on-a-chip (SoC) for epi-retinal and neural prostheses. Taking an AC input, an integrated power telemetry circuits is capable of generating multiple DC voltages with a voltage conversion efficiency of 83% at a load of 100 mW without external diodes or separate power integrated circuits, reducing the form factor of the prosthetic device. A wireless DPSK receiver with a novel noise reduction scheme supports a data rate of 2 Mb/s at a bit-error-rate of 2 ×10⁻⁷. The 1024-channel stimulator array meets an output compliance voltage of ±10 V and provides flexible stimulation waveforms. Through chip-clustering, the stimulator array can be further expanded to 4096 channels. This SoC is designed and fabricated in TSMC 0.18 μm high-voltage 32 V CMOS process and occupies a chip area of 5.7 mm × 6.6 mm. Using this SoC, a retinal implant bench-top test system is set up with real-time visual verification. In-vitro experiment conducted in artificial vitreous humor is designed and set-up to investigate stimulation waveforms for better visual resolution. In our in-vivo experiment, a hind-limb paralyzed rat with spinal cord transection and implanted chronic epidural electrodes has been shown to regain stepping and standing abilities using stimulus provided by the SoC.
No takes yet. Share an insight, caveat, or question.
Lo et al. (2013) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: