Answers are grounded in this paper and cite their sources.
A multiple-disk centrifugal pump based on the Tesla Turbine design has been modified for potential use as an artificial ventricle or ventricular assist device. The pump consists of a series of interconnected parallel disks placed within a spiral volute housing. This pump normally operates as a continuous flow device; however, a controller circuit has been developed to also allow for pulsatile operation. Frequency, systolic duration, systolic rise time, and diastolic decay time can be independently controlled to produce a wide range of pulsatile pressures and flows. This pumping system was tested in vitro on a mock circulatory system using a blood analogue. Inlet and outlet pressures, outlet flow, and motor rotations per minute were continually monitored over a wide range of physiologic operating conditions. The disk pump output was compared with that of other artificial ventricles and produced favorable results. Direct experimental comparisons were made with a Harvard Apparatus pulsatile piston pump. Unlike the Harvard pump, the disk pump does not use valves. Rather, a slight forward rotation of the disks is used to offset the adverse diastolic pressure gradient, which avoids backflow through the device.
No takes yet. Share an insight, caveat, or question.
Miller et al. (1993) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: