Answers are grounded in this paper and cite their sources.
AIMS: Hydroxyapatite coatings for uncemented fixation in total knee arthroplasty can theoretically provide a long-lasting biological interface with the host bone. The objective of this study was to test this hypothesis with propriety hydroxyapatite, peri-apatite, coated tibial components using component migration measured with radiostereometric analysis over two years as an indicator of long-term fixation. PATIENTS AND METHODS: A total of 29 patients at two centres received uncemented PA-coated tibial components and were followed for two years with radiostereometric analysis exams to quantify the migration of the component. RESULTS: While there was significant variation in individual migration patterns, the overall migration of the tibial component in the study group demonstrated a pattern of initial migration followed by stabilisation after one year, with mean maximum total point motion (MTPM) of 0.02 mm (standard deviation (sd) 0.20) between one and two years post-operatively. The direction of greatest motion was subsidence, which stabilised at three months post-operatively (mean translation of 0.21 mm, sd 0.40). CONCLUSION: 2017;99-B:1596-1602.
No takes yet. Share an insight, caveat, or question.
Dunbar et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: