The description of simple and fast calibration procedures used for double-probe triaxial fluxgate gradiometer is provided in this paper. The calibration procedure consists of three basic steps. In the first step both probes are calibrated independently in order to reach constant total field reading in every position. Both probes are numerically aligned in the second step in order that the gradient reading is zero in homogenous magnetic field. The third step consists of periodic drift calibration during measurement. The results and detailed description of each calibration step are presented and discussed in the paper. The gradiometer is finally verified during the detection of the metal object in the measuring grid.