We developed a dual-wavelength laser speckle imaging system using a single industrial-grade color CCD camera with Bayer filters to simultaneously image changes in blood flow, blood volume, and oxygenation. One frame of a color image recorded with dual-wavelength laser illumination provides not only the intensity fluctuation of the speckle pattern, but also the dual-wavelength optical reflectance signal. The method was validated using a tissue phantom and cuff ischemia experiments in the human arm. This system achieves complete time synchronization, unlike conventional time-sharing systems. Compared with a multicamera system, it also avoids the problem of image registration and can be less expensive.