This study compared the effectiveness of pulsed tones (200 ms on/200 ms off) and continuous tones (1 to 2 s) for measuring thresholds in patients with tinnitus. Sixty veterans with tonal tinnitus and a history consistent with noise-induced hearing loss participated. Half were assigned to a pulsed-tone group; half were assigned to a continuous-tone group. Air-conduction thresholds were measured twice for each patient in a single session for octave intervals between 250 Hz and 4000 Hz. Results indicated no significant difference between groups for test-retest differences, but overall more presentations were required to measure thresholds with continuous tones than with pulsed tones at 4000 Hz. Moreover, the continuous-tone group had, on average, more false-positive responses than the pulsed-tone group. This study supports the clinical belief that pulsed tones are preferred over continuous tones for measuring thresholds for persons with tinnitus.
No takes yet. Share an insight, caveat, or question.
Mineau et al. (1997) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: