Key result
Durata ICD lead linked to ~82% lower annual electrical failure vs Riata.
Why the study?
Suspicions regarding the Durata defibrillator lead's failure rates arose due to design similarities and a common manufacturer with the recalled Riata lead.
Does the Durata defibrillator lead have better failure-free survival compared to Riata and Sprint Quattro leads in patients requiring ICDs?
Cohort (n=2,475)
No
Does the Durata defibrillator lead have better failure-free survival compared to Riata and Sprint Quattro leads in patients requiring ICDs?
The Durata defibrillator lead demonstrates excellent failure-free survival, comparable to the Sprint Quattro and significantly better than the recalled 8 Fr Riata lead.
Answers are grounded in this paper and cite their sources.
Durata may be preferable to recalled Riata leads in ICD patients; leaves open need for randomized durability confirmation.
AIMS: Given design similarities and a common manufacturer, there have been suspicions regarding the Durata™ defibrillator (ICD) lead, in the aftermath of the Riata™ class I recall. We therefore examined the failure-free survival rates of the Durata™ compared with the Riata™ and Sprint Quattro™ ICD leads. METHODS AND RESULTS: All patients (n = 2475) implanted with a Durata (n = 828), Riata [n = 627; 8 Fr. (n = 472) and 7 Fr. (n = 155)], or Sprint Quattro (n = 1020) leads at our institution were included and Kaplan-Meier failure-free survival curves were constructed for all leads. Lead failure was defined as electrical malfunction resulting in lead replacement, excluding dislodgements or perforations. Annual electrical failure rates were 0.3%, 1.7, and 0.3% for the Durata, Riata, and Sprint Quattro leads, respectively (P < 0.0001 for the comparison of Durata to Riata and P = 0.1.0 for the comparison of Durata to Sprint Quattro). The failure-free survival of the Durata lead was significantly better than that of the Riata lead (P < 0.0001) and similar to that of the Sprint Quattro (P = 0.94). The 7 Fr. Riata ST lead had better survival compared with the 8 Fr. Riata lead (P = 0.050) and comparable survival with the Durata lead (P = 0.12). CONCLUSION: The Durata lead failure-free survival is significantly better than the 8 Fr. Riata, albeit at a shorter follow-up time. Riata and comparable with that of the 7 Fr. Riata ST and the Sprint Quattro ICD leads. These data provide an insight into the mechanism of electrical failure of Riata leads and have implications for patient management.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2013) conducted a cohort in Patients requiring an implantable cardioverter-defibrillator (ICD) (n=2,475). Durata ICD lead vs. Riata and Sprint Quattro ICD leads was evaluated on Annual electrical failure rate (electrical malfunction resulting in lead replacement, excluding dislodgements or perforations) (p=<0.0001). The Durata ICD lead demonstrated an annual electrical failure rate of 0.3%, which was significantly lower than the Riata lead (1.7%, P<0.0001) and similar to the Sprint Quattro lead.