Answers are grounded in this paper and cite their sources.
Capillary-based TEER measurements are affected by parasitic capacitances that distort the impedance spectrum and compromise reliable determination of barrier resistance of 3D cell structures, e.g. organoids. In this study, an actively shielded glass capillary concept is introduced to suppress parasitic coupling effects in four-electrode impedance measurements. Circuit simulations and experimental pre-tests demonstrate that capacitive artifacts significantly influence unshielded configurations, whereas active shielding stabilizes the impedance response within the frequency range relevant for TEER analysis.
No takes yet. Share an insight, caveat, or question.
Weisgerber et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: