A Ugi multicomponent reaction was employed to synthesize a ferrocene (Fc)-branched polyethyleneimine derivative. This polymer was evaluated as a redox-active matrix for the fabrication of an electroanalytical disposable glucose biosensor based on a carbon PE-printed electrode functionalized with glucose oxidase as a biorecognition element. The resulting biosensor displayed a linear response up to 35 mM glucose at pH 6.5, with a limit of detection of 33 µM. The enzyme biosensor reached 95% of the steady-state current within approximately 50 s and showed a sensitivity of 0.92 µA/mM. In addition, this analytical device exhibited high selectivity, good reproducibility, and excellent storage stability, retaining strong analytical performance after 45 days at 4 °C. The proposed electrochemical biosensor was successfully applied to glucose determination in red and white wine samples.
No takes yet. Share an insight, caveat, or question.
Urraca et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: