In order to improve and extend the application of U- and Th-series isotope measurements a new technique for MIC-ICP-MS (multiple ion counting inductively coupled plasma mass spectrometry) was developed. This method uses either two (double MIC) or three (triple MIC) multipliers to measure in static mode in turn the U-isotopes 233U, 234U, 235U and 236U. For online internal standardization, mass bias correction and cross calibration of the multipliers a 233U/236U double spike was used. Applying this method the level of 1‰ 2σ internal precision is reached in less than 30 min, consuming less than 5 ng of total U. The multi-static MIC-ICP-MS method improves the precision by a factor of 5 for U isotope measurement compared with TIMS and by a factor of 2 compared with published ICP-MS methods. Repeated measurements of uranium CRM-145 (otherwise NBL 112A) were performed to test the reproducibility and accuracy of the method. An average 234U/238U value of (5.28584 ± 0.00087) × 10−5 or a δ234U value of −36.96 ± 0.16, respectively, was determined for CRM-145 in good agreement with reference data. Based on the precisely determined U isotope ratios we furthermore developed a combined multi-static U- and Th-measurement method. This approach extends this new MIC-ICP-MS technique to those U- and Th-series isotope systems having no constant isotope ratio for normalization (e.g., 230Th/232Th, 230Th/229Th) and is the key for new approaches to determining 226Ra/228Ra and 231Pa/233Pa. The applicability of the methods presented here is demonstrated by accurate dating of very young corals (10–350 y).
No takes yet. Share an insight, caveat, or question.
Fietzke et al. (2005) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: