Key result
Hybrid vHPSD/HPSD ablation matches cryoballoon PVI efficacy while cutting fluoroscopy time by ~67%.
Why the study?
A hybrid approach with very high-power short-duration posteriorly and ablation-index guided HPSD anteriorly may optimize efficiency and effectiveness for point-by-point pulmonary vein isolation compared to cryoballoon ablation in symptomatic AF patients.
Does hybrid vHPSD/HPSD radiofrequency ablation improve procedural efficiency and safety compared to cryoballoon ablation in patients with symptomatic atrial fibrillation?
Cohort (n=110)
No
Does hybrid vHPSD/HPSD radiofrequency ablation improve procedural efficiency and safety compared to cryoballoon ablation in patients with symptomatic atrial fibrillation?
A hybrid vHPSD/HPSD radiofrequency ablation strategy for pulmonary vein isolation is as fast and effective as cryoballoon ablation, while significantly reducing radiation exposure.
Answers are grounded in this paper and cite their sources.
May reduce fluoroscopy in AF ablation; hypothesis-generating and requires RCTs before practice change.
Introduction A hybrid approach with very high-power short-duration (vHPSD) posteriorly and ablation-index guided HPSD (50 W) anteriorly seems to be an optimal balance between efficiency and effectiveness for point-by-point pulmonary vein isolation (PVI). The aim of the current study is to compare vHPSD/HPSD ablation to cryoballoon ablation (CBA) in patients with symptomatic atrial fibrillation (AF). Methods and results In this retrospective single-center study, we identified 110 consecutive patients who underwent their first PVI with either vHPSD/HPSD (n = 54) or CBA (n = 56). We compared procedural efficacy, efficiency, safety, and long-term outcomes. Baseline characteristics of both groups were comparable; however, patients in the vHPSD/HPSD group had larger left atrial volume index (35, IQR 27–45 vs. 28, IQR 21–36 ml/m2, P = 0.005). Complete PVI was achieved in all patients except two CBA cases (100% vs. 96.4%, P = 0.50). First-pass isolation rate was 79.6% in the hybrid group. Procedure times were similar between groups (53, IQR 47–63 vs. 55, IQR 49–65 min, P = 0.35), but fluoroscopy time was shorter in the vHPSD/HPSD group (3.9 [2.7, 5.6] vs. 11.9 [9.3, 14.9] min, P < 0.001). There were 3 temporary phrenic nerve palsies (5.4%) in the CBA group which resolved within 1 year. The 1-year freedom from any atrial tachyarrhythmias after a single procedure was similar between groups (68.5% vs. 73.2%, P = 0.56). During repeat procedure, the durability of PVI was comparable. Conclusions The use of vHPSD/HPSD ablation renders point-by-point PVI as fast and effective as CBA. Furthermore, it has lower radiation exposure compared to CBA.
No takes yet. Share an insight, caveat, or question.
Hoogendijk et al. (2025) conducted a cohort in Symptomatic atrial fibrillation (n=110). Hybrid very high-power short-duration (vHPSD) and high-power short-duration (HPSD) radiofrequency ablation vs. Cryoballoon ablation (CBA) was evaluated on Complete pulmonary vein isolation (p=0.50). Hybrid vHPSD/HPSD ablation achieved similar complete pulmonary vein isolation rates (100% vs 96.4%) and procedure times (53 vs 55 min) compared to cryoballoon ablation, but significantly reduced fluoroscopy time (3.9 vs 11.9 min).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: