Answers are grounded in this paper and cite their sources.
We have evaluated an immunoturbidimetric method for the estimation of urinary albumin. The method, besides being easy to perform and cost-effective, was sensitive enough to detect an even slightly increased albumin excretion (detection limit 5 mg/l). Within-run reproducibility was 1.8 and 2.1%, and between-run reproducibility 2.9 and 4.3% in samples containing 16.1-17.8 mg/l and 50.6-54.0 mg/l of albumin, respectively. The recovery of albumin added to the samples was 98.6-106.6%. Results obtained by this method correlated well with the results obtained by radial immunodiffusion (r = 0.980, n = 44) and radioimmunoassay (r = 0.982, n = 41). The immunoturbidimetric method can be easily adapted for several clinical chemistry analysers.
No takes yet. Share an insight, caveat, or question.
Paloheimo et al. (1987) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: