Answers are grounded in this paper and cite their sources.
A capillary system formed by combining 25 and 100 μm id capillaries was used in the short-end injection mode to determine creatinine and uric acid in human urine. The separation was performed at an electric field intensity of 2.3 kV/cm. Creatinine was determined in a BGE with a composition of 20 mM citric acid/NaOH (pH 3.0), and uric acid was determined in 20 mM MES/NaOH (pH 6.0). Under these conditions, migration times of 12.2 s for creatinine and 8.6 s for uric acid were achieved. The LOD value is 2.4 mg/L for creatinine and 0.9 mg/L for uric acid; the RSD for the migration time varies in the range 0.7-1.1% (intra day) to 1.0-7.5% (inter day); RSDs for the peak areas equalled 3.4-4.0% (intra day) and 4.3-4.7% (inter day). The determined creatinine values in seven urine samples vary in the range 221-1394 mg/L for creatinine and 87-615 mg/L for uric acid. t-Test did not reveal any statistically significant difference between the developed CE methodologies and reference methods - Jaffé reaction for creatinine and enzymatic uricase test for uric acid.
No takes yet. Share an insight, caveat, or question.
Pavlíček et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: