The National High Magnetic Field Laboratory (NHMFL) has completed testing and commissioning of a unique ultrahigh field magnet that provides 36 T in a 32 mm bore with field inhomogeneity and stability better than 1 ppm over a cylinder of 1 cm diameter and length and a duration up to a few hours. While the magnet meets all its primary performance goals, there were some unexpected observations during the testing and commissioning activities. Primarily, while the magnet reaches full current and field, we are unable to charge the magnet as quickly as was anticipated. Also, the magnet occasionally quenches for no apparent reason. Charging the magnet at slower rates reduces the probability of quenches.
No takes yet. Share an insight, caveat, or question.
Bird et al. (2019) studied this question.