OBJECT: Continuous intracranial pressure (ICP) monitoring using an epidural sensor is a common technique used in selected neurosurgical patients. The aim of this study was to assess the safety and accuracy of the Neurodur-P epidural sensor in the clinical setting. METHODS: The zero drift, as well as the medical and technical complications, of using the Neurodur-P sensor placed in the epidural space was evaluated in 106 patients with hydrocephalus of varying causes or with suspected intracranial hypertension. RESULTS: The median duration of ICP monitoring was 8 days (interquartile range [IQR] 6-12 days). In 78 (73.6%) of the 106 patients the pressure reading was recorded at sensor removal. No zero drift was observed in 28 sensors. The median drift was 0 mm Hg with an IQR of -1 to 1 mm Hg. No significant differences were found between patients monitored for < or = 5 days and those monitored for > 5 days (t = 535, p = 0.100). No correlation was found between zero drift and monitoring time (r = 0.153, p = 0.181). Of the 83 patients with a follow-up computed tomography scan, 3 showed a < 1 ml collection of blood at the catheter tip. No clinical infections could be attributed to the devices. Only 1 sensor malfunctioned. CONCLUSIONS: Continuous ICP monitoring using the Neurodur-P sensor is safe, reliable, and easy to perform. At present, using this device is the authors' standard method for the long-term monitoring of patients with alterations in complex cerebrospinal fluid dynamics or with implanted shunts.
No takes yet. Share an insight, caveat, or question.
Poca et al. (2008) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: