Key result
Micra AV2 algorithms more than double the rate of high AV synchrony vs. Micra AV.
Why the study?
Early clinical experience with Micra AV leadless pacemakers showed suboptimal AV synchrony due to inadequate device parameter programming, motivating algorithm enhancements in Micra AV2 to improve automatic AV synchrony and reduce manual programming burden.
Do the enhanced algorithms in the Micra AV2 leadless pacemaker improve atrioventricular synchrony and reduce programming burden compared to the original Micra AV algorithms in virtual patients with complete AV block?
Do the enhanced algorithms in the Micra AV2 leadless pacemaker improve atrioventricular synchrony and reduce programming burden compared to the original Micra AV algorithms in virtual patients with complete AV block?
Absolute Event Rate: 90% vs 43%
p-value: p=<0.001
Simulation data suggests that algorithm updates in the Micra AV2 leadless pacemaker significantly improve automatic atrioventricular synchrony at higher sinus rates and reduce device programming time.
May lessen manual programming burden in leadless pacing; leaves open real-world AV synchrony gains.
PURPOSE: To assess the impact of enhancements to the Atrioventricular Synchrony (AVS) algorithms of a next generation Micra leadless pacemaker (Micra AV2). METHODS: Accelerometer data were extracted from the AccelAV clinical study and were used to create virtual patients. A series of Monte Carlo simulations were run for each virtual patient to compare an enhanced Atrial Sensing Setup algorithm and Auto + A3 Threshold algorithm vs. original algorithms. A real-world survey was also conducted to observe clinical time savings from AVS programming burden reduction. RESULTS: The enhanced Atrial Sensing Setup in Micra AV2 devices demonstrated > 70% AVS in 27 of 30 (90%) patients while 13 of 30 (43%) Micra AV patients had > 70% AVS (p < 0.001) with no manual programming. The Micra AV2 Auto + A3 Threshold without additional manual programming demonstrated improved overall ambulatory AVS in the 80-100 bpm range (84.1%). Based upon survey results, the enhanced Atrial Sensing Setup algorithm accounted for an estimated reduction in median device check time of 13 min per patient. CONCLUSIONS: Simulation-based analyses of the Micra AV2 leadless pacemaker projected significant improvements in automatic AVS at high sinus rates and an increase in the number of patients with > 70% AVS without clinician programming. Real-world survey results reported a reduction in device check time with the improvements. SIGNIFICANCE: Improvements in the AVS algorithms in Micra AV2 allow for better automatic AVS at higher heart rates and reduced clinic utilization burden.
No takes yet. Share an insight, caveat, or question.
Leal et al. (2025) studied Complete atrioventricular block with normal sinus function (n=30). Micra AV2 Atrial Sensing Setup and Auto + A3 Threshold algorithms vs. Micra AV original algorithms was evaluated on Proportion of patients achieving > 70% atrioventricular synchrony (AVS) following Atrial Sensing Setup (p=<0.001). Micra AV2 algorithm enhancements achieved >70% atrioventricular synchrony in 90% of virtual patients compared to 43% with the original Micra AV, without requiring manual programming.