Key result
Early cervical MRI safely performed one day post-pacemaker implant with no device parameter changes.
Why the study?
MRI is generally recommended only after 4-8 weeks of pacemaker implantation to allow healing, but the safety of earlier scanning remains uncertain.
Case Report (n=1)
Urgent MRI may be feasible 1 day after pacemaker implantation; leaves open broader early-scan safety pending prospective data.
Recent data showed that patients with electrical implanted devices may under certain conditions be scanned safely by magnetic resonance imaging (MRI). The device must have been in place preferably for at least 4-8 weeks [Correction added after online publication 31-Aug-2009: number of weeks has been updated.] prior to MR imaging to allow healing and pacemaker pocket formation. We report on a patient with quadriplegia and suspected epidural hematoma referred for MR scan a day after he had a pacemaker implantation. The patient was also pacemaker-dependent. After considering the risk/benefit ratio in this patient, it was decided to perform the scan. The pacemaker was reprogrammed. MRI was performed under strict monitoring. A spinal cord contusion at the level of C1-3 was diagnosed. Based on the imaging findings no invasive procedure was done. Device interrogation found no change in sensing or pacing parameters or in the pacemaker's battery. At the end of the scan, the device was reprogrammed back to the initial settings. In this population, each scan should be discussed thoroughly and the risks to benefit ratio should be considered. Given appropriate precautions, in well-experienced imaging centers, MRI may be safely performed in patients with permanent cardiac electronic implantable devices.
No takes yet. Share an insight, caveat, or question.
Goldsher et al. (2009) conducted a case report in Quadriplegia and suspected epidural hematoma in a pacemaker-dependent patient (n=1). Magnetic resonance imaging (MRI) scan was evaluated on Change in sensing or pacing parameters or pacemaker battery. Cervical MRI was safely performed in a pacemaker-dependent patient one day after implantation, resulting in no changes to sensing, pacing parameters, or battery function.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: