High-field asymmetric waveform ion mobility spectrometry (FAIMS) has been coupled to a quadrupole time-of-flight mass spectrometer for the tandem mass spectrometric analysis of tryptic peptides of pig hemoglobin. Using FAIMS, low levels (fmol/microL) of multiply charged tryptic peptides were separated from relatively intense chemical background such that their tandem mass spectra (MS/MS) lacked many background-related fragment ions observed using a conventional ESI-QqTOFMS instrument. Substantial improvements in both first-order and tandem mass spectra were realized while maintaining approximately the same absolute intensities.
No takes yet. Share an insight, caveat, or question.
Barnett et al. (2002) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: