The antimony film microelectrode (SbFME) is presented for measuring trace heavy metal ions in combination with anodic stripping voltammetry. The SbFME was tested in model solutions of 0.01 M HCl containing Cd(II) and Pb(II) in the presence of dissolved oxygen. The microsensor exhibited good linear behavior in the examined concentration range 20 μg L−1 to 100 μg L−1 and LoD of 1.9 and 3.1 μg L−1 for Cd(II) and Pb(II), respectively. The suitability of SbFME for measuring trace levels of Cu(II) was demonstrated using the standard reference material of natural water exploiting the sensor's specific characteristic of low re‐oxidation signal for antimony.
No takes yet. Share an insight, caveat, or question.
Slavec et al. (2010) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: