We have measured the efficiency (tracks per incident neutron) of pure CR-39 for detecting DD and DT neutrons. Neutrons having average energies of 2.9 MeV (DD) and 14.8 MeV (DT) were produced by a 200-keV electrostatic accelerator and the neutron yields were measured using the associated particle counting technique. All CR-39 samples irradiated by DD or DT neutrons were etched for 2 h in a 70°, 6.25-N_ NaOH bath. For bare CR-39, the efficiencies for detecting 2.9- and 14.8-MeV neutrons were found to be (1.3±0.4)×10−4 and (5.0±1.8)×10−5, respectively. We also investigated using CR-39 and polyimide as proton radiators. For detecting 2.9-MeV neutrons, the radiators had no significant effect on efficiency; but for detecting 14.8-MeV neutrons the polyimide radiator increased the efficiency to (7.8±2.8)×10−5.