Wire mesh sensors (WMS) are fast imaging instruments that are used for gas–liquid and liquid–liquid two-phase flow measurements and experimental investigations. Experimental tests were conducted at Helmholtz-Zentrum Dresden-Rossendorf to test both the capacitance and conductance WMS against a gamma densitometer (GD). A small gas–liquid test facility was utilized. This consisted of a vertical round pipe approximately 1 m in length, and 50 mm internal diameter. A 16 × 16 WMS was used with high spatial and temporal resolutions. Air–deionized water was the two-phase mixture. The gas superficial velocity was varied between 0.05 m s −1 and 1.4 m s −1 at two liquid velocities of 0.2 and 0.7 m s −1 . The GD consisted of a collimated source and a collimated detector. The GD was placed on a moving platform close to the plane of wires of the sensor, in order to align it accurately using a counter mechanism, with each of the wires of the WMS, and the platform could scan the full section of the pipe. The WMS was operated as a conductivity WMS for a half-plane with eight wires and as a capacitance WMS for the other half. For the cross-sectional void (time and space averaged), along each wire, there was good agreement between WMS and the GD chordal void fraction near the centre of the pipe.
No takes yet. Share an insight, caveat, or question.
Sharaf et al. (2011) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: