Background and purpose Intermediate-dose spillage is central to normal-brain sparing in stereotactic radiosurgery, yet fixed adverse-event and relative indices do not define achievable plan quality for single-target linear accelerator treatments. The aim was to derive a volume-dependent absolute benchmark for the 50% prescription isodose volume (V 50% ; equivalent to V 10 Gy for 20 Gy plans) for brain metastases. Materials and methods In total, 276 single-isocenter single-target plans were analyzed: 178 idealized model plans and 98 clinical plans. Power-law regression in log-log space modeled spillage against planning target volume (PTV, cm 3 ). The volume-scaling exponents were 0.83 for the ideal references and 0.75 for the simplified clinical threshold. The clinical upper threshold was derived using q = 0.90 quantile regression. Threshold-exceeding plans were reviewed for anatomical or technical justification. Results In the model cohort V 50% value correlated strongly with PTV ( r = 0.994, p < 0.001). The ideal mean model was V 50% = 3.71 × PTV 0.83 , the conservative ideal-planning reference was V 50% = 4.10 × PTV 0.83 , and the clinical upper threshold was V 50% = 5.00 × PTV 0.75 . Ten of 98 clinical plans (10%) exceeded the threshold. Seven lacked anatomical justification and were replanned, reducing V 50% by a median of 14% (range 7–29%) without compromising target coverage. Conclusions A volume-based V 50% benchmark provided a complementary quality-assurance reference for single-target linear accelerator radiosurgery for brain metastases. It flagged plans that warranted structured review alongside target coverage, organs of interest, conformity, and gradient metrics.
No takes yet. Share an insight, caveat, or question.
Nour et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: